10 Sensors
900 Touch Control
■■■■■■■■
67
Calibration interval (for pH electrodes, ISE electrodes and conductivity
measuring cells only)
Working life
etc.
NOTE
If data is read out from the data chip of an intelligent sensor, then a
check is made whether the sensor list already contains a sensor of the
same serial number. If this is the case, then the older data set will
always be overwritten by the new data set, no matter whether the
data set in the sensor list or the data set on the data chip is the most
recent one.
[New]
Add a new sensor to the list (see Chapter 10.1, page 67).
[Delete]
Delete the selected sensor from the list.
[Edit]
Edit the data of the selected sensor (see Chapter 10.2, page 68).
10.1 Adding a new sensor
Before you can use a sensor, you must add it to the sensor list. To do this,
use the button [New].
Conventional sensors:
The properties dialog is opened after the sensor type has been
selected, see following chapter. The following sensor types can be
selected:
pH electrode
Metal electrode (Pt electrode, Ag Titrode, Ag/AgCl reference
electrode, etc.)
Ion-selective electrode
Other sensor, e.g. Spectrosense
Temperature sensor
Conductivity measuring cell
Intelligent sensors (also known as iTrodes):
If the 854 iConnect with iTrode is connected to a control device, then
the sensor is automatically entered in the sensor list and can be config-
ured, see following chapter.