13.1 Automatic system test
■■■■■■■■■■■■■■■■■■■■■■
128
■■■■■■■■
900 Touch Control
[GLP tests HW/SW]
Document the GLP tests "Measurement" and "Titration" (see Chapter
13.3, page 129).
[System validation]
Document the system validation (see Chapter 13.4, page 133).
[Monitoring]
Define time intervals for system maintenance and backups (see Chapter
13.5, page 137).
[View test data]
Display the data of the selected test.
13.1 Automatic system test
The system test is carried out automatically when the Touch Control is
switched on. The result of each individual test is shown in green if no
error has occurred. If a result is shown in red, then an error occurred
when the respective test was carried out. Switch the Touch Control off
and back on again. If the error still occurs pleased notify Metrohm Service.
Print report at system start
on | off (Default value: off)
If this parameter is activated, then the result of the system test will be
printed out automatically when the Touch Control is switched on.
13.2 Test tools
Main dialog: System GLP manager Test tools
You can create a list of test tools to be used in the tests in the GLP man-
ager.
The following test tools have already been defined:
767 Calibrated Reference: Device for checking measuring inputs
and electrode cable.
822 Titr.Curve Simulator: Curve simulator for checking the hard-
ware and software.
773 pH/mV Simulator: Device for checking measuring inputs and
electrode cable.
868 UR Generator: Device for checking measuring inputs and elec-
trode cable.
[New]
Add a new test tool to the list.