16.5 Method options
■■■■■■■■■■■■■■■■■■■■■■
178
■■■■■■■■
900 Touch Control
Upper limit
Input range –999999999 - 9999999999
Default value 9999999999
Display message
on | off (Default value: off)
If this parameter is activated, a message is displayed in case the limits are
infringed. You can select whether you want to continue with the determi-
nation or cancel the run.
If this parameter is deactivated, then the message will be entered only in
the message list of the determination.
Timeout
The message is displayed during this time. Afterwards, the determination
is continued automatically.
Input range 0 - 999999 s
Default value 30 s
Selection off
off
The determination will not be continued until after the message has
been confirmed.
16.5.4 Method properties
Main dialog: Edit parameters Method options Properties
Detailed information concerning the method is displayed in the dialog
Method options / Properties.
Method status
Current status of the method.
new
The method has been newly created and not yet saved.
saved
The current method has been saved.
modified
The current method has been modified.
reviewed
The current method has been signed at level 1.
released
The current method has been signed at level 2.
Created by
User who created the method.
Created on
Date and time at which the method was created.