29 Parameters
900 Touch Control
■■■■■■■■
337
Extraction time
Minimum duration of the titration. The titration will not be stopped during
the extraction time, even if the endpoint has already been reached. The
titration is however stopped if a stop condition is fulfilled during this time
(see Chapter 29.5.3, page 337). Entering an extraction time may e.g. be
advisable if a Karl Fischer oven is used.
Input range 0 - 999999 s
Default value 0 s
Temperature
Temperature entered manually. The temperature is being continuously
measured when a temperature sensor is connected and when Temp.
measurement = automatic or continuous is defined (see sensor dialog
of the command).
Input range –20.0 - 150.0 °C
Default value 25.0 °C
Time interval MP
Time interval for entering a measuring point in the measuring point list.
The measuring point list is limited to 1000 measuring points.
Input range 0.1 - 999999.0 s
Default value 2.0 s
29.5.3 Stop conditions
The conditions for canceling the titration are defined under [Stop condi-
tions], if this does not occur automatically. This could be the case when
the endpoint set is not reached or if the stop criterion (see "Stop crite-
rion", page 335) is not fulfilled.
Stop time
The titration is stopped when the specified time has elapsed since the start
of the titration.
Input range 1 - 999999 s
Selection off
Default value off
29.5.4 Conditioning
The conditions required for conditioning are defined under [Condition-
ing].