Page view 129
11.9 USB/RS-232 adapter
■■■■■■■■■■■■■■■■■■■■■■
114
■■■■■■■■
900 Touch Control
11.9 USB/RS-232 adapter
List of devices: USB/RS-232 adapter Edit COM interface Edit
As a rule, balances are equipped with a serial RS-232 interface. To con-
nect a balance, you will require the 6.2148.050 cable. You can adjust the
interface parameters in the dialog Edit device / Port parameters. These
settings must match the settings on the connected device.
NOTE
If you make changes to parameter settings in this dialog, then you must
switch the Touch Control off and back on in order for the changes to
take effect.
Baud rate
Transfer rate in characters per second.
Selection 1200 | 2400 | 4800 | 9600 | 19200 | 38400 |
57600 | 115200
Default value 9600
Data bits
Number of data bits.
Selection 7 | 8
Default value 8
Parity
Type of parity testing.