29 Parameters
900 Touch Control
■■■■■■■■
299
Titration rate
slow optimal fast
Signal drift
– pH, U and Ipol
– Upol
20.0 mV/min
20.0 µA/min
50.0 mV/min
50.0 µA/min
80.0 mV/min
80.0 µA/min
Min. waiting time 0 s 0 s 0 s
Max. waiting time 38 s 26 s 21 s
29.2.3 Stop conditions
The conditions for canceling the titration are defined under [Stop condi-
tions].
Stop volume
The titration is stopped when the specified volume has been dosed since
the start of the titration. This volume should be adjusted to the size of the
titration vessel in order to prevent the contents from running over.
Input range 0.00000 - 9999.99 mL
Default value 100,000 mL
Selection off
Stop meas. value
The titration is canceled when the specified measured value has been
reached since the start of the titration.
The values that are actually measurable may vary from the values specified
here, depending on the hardware. The binding measurement value ranges
can be found in the "Technical specifications" chapter for the control
device used.
Measuring mode pH:
Input range –20.000 - 20.000
Selection off
Default value off
Measuring mode U, Ipol:
Input range
–2,000.0 - 2,000.0 mV
Selection off
Default value off
Measuring mode Upol:
Input range
–200.0 - 200.0 µA
Selection off
Default value off