29 Parameters
900 Touch Control
■■■■■■■■
367
Sensor
Selection of the sensor from the sensor list. The selection depends on the
measuring mode. Sensors are defined under System Sensors. You can
also enter a sensor name which is not contained in the sensor list. When a
determination is started there is a check whether the sensor is contained
in the sensor list.
Selection Selection of configured sensors
Temp. measurement
Type of temperature measurement.
Selection continuous | automatic | off
Default value automatic
continuous
A temperature sensor must be connected. The temperature is mea-
sured continuously.
automatic
If a temperature sensor is connected then the temperature will be
measured continuously. Otherwise, the temperature entered manually
will be used (see dialog of the titration and measuring parameters).
off
The temperature will not be measured. The temperature entered man-
ually is used (see dialog of the titration and measuring parameters).
29.7.8 Dosing device
The parameters for the dosing device are edited under [Dosing device].
Dosing device
Selection of the MSB connector the dosing device is connected to. Every
MSB connector will always be displayed.
Selection 1 | 2 | 3 | 4
Default value 1
Titrant
Selection of the titrant from the titrant list. We recommend always select-
ing the titrant. This ensures that the correct data (titer, concentration, etc.)
is always used for the calculation. Titrants are defined under Sys-
tem Titrants. You can also enter a name which is not contained in the
titrant list. When a determination is started there is a check whether the
solution is contained in the list.
For buret units with integrated data chip, a check is made in the method
run to verify whether the correct titrant has been attached and whether
the type of dosing drive matches. For buret units without integrated data