24 Results and more determination data
900 Touch Control
■■■■■■■■
225
24.6 Saving a determination
Results dialog: More data ▶ Load/ Save ▶ Save
Determinations can be saved as follows:
■ Manually in this dialog.
The suggested default setting for file names is the first 16 characters of
Identification 1, followed by the determination time in the format
YYYYMMDD-hhmmss. Extending the file name with the determination
time has the advantage that the file names of all determinations are
unique.
■ Automatically at the end of the determination (see Chapter 16.5.6,
page 180).
The following data is stored for each determination:
■ Calculated results
■ Measuring point lists, equivalence points, curves and variables which
were generated during the determination
■ Determination properties
■ Statistics (if carried out)
■ Method with which the determination was carried out
■ Common variables (values and properties of all common variables at
the time of the determination)
■ Titrant data of the titrant used
■ Calibration data of the sensor used
24.7 Curves
Results dialog: Curve
The curve of the current determination is displayed in the Results / Curve
dialog. You can configure the curve display.