29.8 Manual titrations (MAT)
■■■■■■■■■■■■■■■■■■■■■■
372
■■■■■■■■
900 Touch Control
Filling rate
Rate at which the dosing cylinder is filled.
Input range 0.01 - 166.00 mL/min
Selection maximum
Default value maximum
maximum
The maximum filling rate depends on the cylinder volume (see Chapter
32, page 488).
Temperature
Temperature entered manually. The temperature is being continuously
measured when a temperature sensor is connected and when Temp.
measurement = automatic or continuous is defined (see sensor dialog
of the command). This value is used for temperature correction in pH
measurements.
Input range –20.0 - 150.0 °C
Default value 25.0 °C
29.8.2 Control device
The control device the determination is carried out with is selected under
[Control device]. Control devices are defined under System Device
manager.
NOTE
This button is only displayed when several control devices have been
configured.
Control device
Selection of the control device from the list of devices. Only those devices
are displayed which are able to carry out the command. The control devi-
ces do not have to be connected. This allows to transfer methods more
easily from one system to another.
Selection Selection of configured control devices
29.8.3 Sensor
The parameters for the sensor are edited under [Sensor].