5.9 Servicing the inline filter
■■■■■■■■■■■■■■■■■■■■■■
76
■■■■■■■■
930 Compact IC Flex Oven/SeS/Deg (2.930.2460)
Inserting a new filter
1
Inserting a new filter
■ Place the new filter flat in the filter housing (22-1) using tweezers.
2
Mounting the filter screw
■ Screw the filter screw (22-2) back into the filter housing (22-1)
and tighten by hand. Then use two adjustable wrenches
(6.2621.000) to tighten it slightly.
3
Remounting the inline filter
■ Screw pressure screws back onto the inline filter.
4
Rinsing the inline filter
■ Dismantle the guard column (if present) and the separation col-
umn and replace with a coupling (6.2744.040).
■ Rinse the instrument with eluent.
■ Reinsert the columns after 10 minutes.