930 Compact IC Flex Oven/SeS/Deg (2.930.2460)
Connecting the separation column
1
Removing the stoppers
CAUTION
When inserting the column, ensure that it is inserted correctly
based on the marked flow direction.
■ Remove the stoppers from the separation column.
2
Installing the input of the separation column
There are 3 possibilities:
■ Attaching the bottom end of the separation column directly to
the guard column.
or
■ If the guard column is connected to the separation column using
a provided connection capillary, connect the bottom end of the
separation column to the guard column's output capillary using
the provided PEEK pressure screw (6.2744.070).
or
■ If no guard column is used (not recommended), connect the col-
umn input capillary to the input of the separation column using a
short pressure screw (6.2744.070).
3
Rinsing the separation column
■ Place a beaker under the separation column's outlet.
■ Start manual control in MagIC Net and select the high-pressure
pump: Manual ▶ Manual control ▶ Pump
– Flow: in accordance with column leaflet
– On
■ Rinse the separation column with eluent for approximately 10
minutes.
■ Use manual control in MagIC Net to stop the high-pressure pump
again: Off.
4
Removing the coupling
■ Remove the coupling (6.2744.040) from the column output capil-
lary.