2.3 Feed-throughs for capillaries and cables
■■■■■■■■■■■■■■■■■■■■■■
10
■■■■■■■■
930 Compact IC Flex Oven/SeS/Deg (2.930.2460)
2.3 Feed-throughs for capillaries and cables
Multiple openings are available for leading capillaries into the instrument
and for leading capillaries and cables out of the instrument:
■ Openings on the door
■ Openings on the back panel
■ Ducts between the instrument and the base tray as well as between
the instrument and the bottle holder (see Figure 5, page 12)
Openings on the door
Figure 3
Feed-throughs on the door
1
Luer connector
For connecting a capillary from inside and
for inserting a syringe (6.2816.020) from
outside. For manual sample injection.
2
Opening for capillaries
For up to 3 capillaries.
An opening for up to 3 capillaries is located on the door of the instru-
ment.
The two Luer connections above are not actually openings; the capillaries
are fastened to the Luer connection from within using PEEK pressure
screws. You can use a syringe to inject or draw out liquid from the out-
side.