3 Installation
930 Compact IC Flex Oven/SeS/Deg (2.930.2460)
■■■■■■■■
45
MagIC Net detects the instrument and all of its modules.
3
Starting equilibration
■ Start the equilibration in MagIC Net: Workplace ▶ Run ▶ Equi-
libration ▶ Start HW.
4
Deaerating the high-pressure pump
■ Push the end of the purging needle (6.2816.040) over the end of
the deaerating capillary on the purge valve.
■ Insert the syringe (6.2816.020) in the Luer connector of the purg-
ing needle.
■ Open the purge valve using the rotary knob (approx. ½ turn).
■ Switch on the high-pressure pump in MagIC Net.
■ Use the syringe to aspirate eluent until there are no more air bub-
bles in the aspirated eluent.
■ Switch off the high-pressure pump in MagIC Net.
■ Seal the purge valve using the rotary knob.
■ Remove the syringe from the purging needle.
■ Pull the purging needle out of the deaerating capillary.
5
Rinsing the instrument without columns
■ Rinse the instrument (without columns) with eluent for 5 minutes.