Metrohm 930 Compact IC Flex Oven/SeS/Deg User Manual Page 24

  • Download
  • Add to my manuals
  • Print
  • Page
    / 113
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 23
3.3 Removing transport locking screws
■■■■■■■■■■■■■■■■■■■■■■
16
■■■■■■■■
930 Compact IC Flex Oven/SeS/Deg (2.930.2460)
3.3 Removing transport locking screws
To avoid damage to the drives for the high-pressure pump and the vac-
uum pump during transport, the pumps are secured with transport lock-
ing screws. These are located at the rear of the instrument and labeled
with Transport security screws.
Remove these transport locking screws before the initial start-up.
Accessories For this step you need:
Hex key 4 mm (6.2621.030)
Remove the transport locking screws
Figure 6
Removing the transport locking screws
1
Transport locking screws
For the vacuum pump. Only present if the
instrument has a degasser.
2
Transport locking screws
For the high-pressure pump.
1
Remove all of the transport locking screws with the hex key.
Store the transport locking screws in a safe place. Reinsert the transport
locking screws each time you transport the instrument a significant dis-
tance.