3 Installation
930 Compact IC Flex Oven/SeS/Deg (2.930.2460)
■■■■■■■■
27
Figure 12 Exchanging the sample loop
1
Pressure screw
Fastened to Port 6.
2
Sample loop
3
Pressure screw
Fastened to Port 3.
Exchanging the sample loop
1
Removing the existing sample loop
■ Unscrew the pressure screws (6.2744.010) at Port 3 and Port 6.
■ Remove the sample loop.
2
Installing a new sample loop
■ Fasten one end of the sample loop to Port 3 using a PEEK pres-
sure screw (6.2744.010).
■ Use the second PEEK pressure screw (6.2744.010) to fasten the
other end of the sample loop to Port 6.