3 Installation
930 Compact IC Flex Oven/SeS/Deg (2.930.2460)
■■■■■■■■
47
■ Guard column (suitable for the separation column)
Connecting the guard column
1
Removing the coupling
Remove the coupling installed between the column input capillary
and the column output capillary for the initial start-up.
2
Preparing the guard column
■ Remove the stopper and the sealing cap from the guard column.
3
Connecting the guard column
CAUTION
When inserting the guard column, ensure that it is inserted cor-
rectly based on the marked flow direction (if specified).
■ Fasten the input of the guard column to the column input capil-
lary using a short pressure screw (6.2744.070).
■ Optional: In case the guard column is connected to the separation
column using a provided a connection capillary, fasten the con-
nection capillary to the guard column's output using the pressure
screw also provided.
Rinsing the guard column
1
Rinsing the guard column
■ Place a beaker under the guard column's output.