67
Has the pressure dissipated?
Is the instrument switched off?
1
Release the coupling from the pressure screw and seal it with a stop-
per.
2
Unscrew and remove the pressure screw on the pump head's output
(16-8).
3
Unscrew and remove the four fastening screws (16-3) using the hex
key.
4
Remove the pump head (16-4).
Replacing the piston seal
Carry out the following work on both pistons in turn.
Servicing a piston consists of the following tasks:
1. Replace the piston seal.
2. Optional: Clean or replace the zirconium oxide piston.
3. Reinstall the piston.