3.2 Remote control commands
758 KFD Titrino
90
&SmplData
Object Description Input range Reference
& Root
:
├ SmplData Sample data
: ├ .Status Status of silo memory ON, OFF 3.2.2.92.
├ .OFFSilo Current sample data
│ ├ .Id1 Sample identification 1 up to 8 ASCII char 3.2.2.93.
│ ├ .Id2 Sample identification 2 up to 8 ASCII char ditto
│ ├ .Id3 Sample identification 3 up to 8 ASCII char ditto
│ ├ .ValSmpl Sample size ±X.XXXXX ditto
│ ├ .UnitSmpl Unit of sample size up to 5 ASCII char ditto
├ .ONSilo Current sample data
│ ├ .Counter Counter of silo memory
│ │ ├ .MaxLines Maximum lines read only 3.2.2.94.
│ │ ├ .FirstLine First line read only ditto
│ │ ├ .LastLine Last line read only ditto
│ ├ .EditLine Editing silo lines
│ │ ├ .1 1
st
silo line
│ │ │ ├ .Method Method name up to 8 ASCII char 3.2.2.95.
│ │ │ ├ .Id1 Sample identification 1 up to 8 ASCII char ditto
│ │ │ ├ .Id2 Sample identification 2 up to 8 ASCII char ditto
│ │ │ ├ .Id3 Sample identification 3 up to 8 ASCII char ditto
│ │ │ ├ .ValSmpl Sample size ±X.XXXXX ditto
│ │ │ ├ .UnitSmpl Unit of sample size up to 5 ASCII char ditto
│ │ │ ├ .C24 Value of variable C24 read only ditto
│ │ │ ├ .C25 Value of variable C25 read only ditto
│ │ │ ├ .Mark Mark of silo line read only ditto
│ │ ├ up to 255 lines
│ ├ .DelLine Delete silo line $G 3.2.2.96.
│ │ ├ .LineNum Line number 1...255, OFF ditto
│ ├ .DelAll Delete silo line $G 3.2.2.97.
│ ├ .CycleLines Cycle lines ON, OFF 3.2.2.98.
│ ├ .SaveLines Save results ON, OFF 3.2.2.99.