3.2 Remote control commands
758 KFD Titrino
84
*Parameter Tree part "Parameters for MEAS"
.Measuring Measuring parameters
.SignalDrift Drift for meas.value acquisition depends on meas.quant. 3.2.2.34.
.UnitSigDrift Unit of measured value drift read only ditto
.EquTime Equilibrium time 0...9999, OFF ditto
.MeasInput Measuring input 1, 2, diff. 3.2.2.35.
.Ipol Polarization current 0...1...±127 ditto
.Upol Polarization voltage 0...400...±1270 ditto
.PolElectrTest Test for polarized electrodes ON, OFF ditto
.Temp Titration temperature –170.0...25.0...500.0 3.2.2.36.
.TDelta Time interv.for meas.acquisition 1...2...999999 3.2.2.37.
.Statistics Statistics
.Status Status of statistics calculation ON, OFF 3.2.2.26.
.MeanN No. of individual determinations 2...20 ditto
.ResTab Result table
.Select original,delete n,delete all ditto
.DelN Deletion of individual results 1...20 ditto
.Presel Preselections
.IReq Request of Id's after start id1, id1&2, all, OFF 3.2.2.28.
.SReq Request of sample size after start value, unit, all, OFF ditto
.LimSmplSize Limits for sample size 3.2.2.29
.Status Status of limit control ON, OFF ditto
.LoLim Lower limit 0.0...999 999 ditto
.UpLim Upper limit 0.0...999 999 ditto
.ActPulse Output of a pulse ON, OFF 3.2.2.30.
*Parameter Tree part "Parameters for CAL"
.Calibration Calibration parameters
.MeasInput Measuring input 1, 2, diff. 3.2.2.38.
.CalTemp Calibration temperature –20.0...25.0...120.0 3.2.2.39.
.Buffer
.1
.Value pH value of buffer 1 0...7.00...±20.00 3.2.2.40.
.2
.Value pH value of buffer 2 0...4.00...±20.00, OFF ditto
: up to 9 buffers
.SignalDrift Drift for meas.value acquisition 0.5...2...999, OFF 3.2.2.41.
.EquTime Equilibrium time 0...100...9999, OFF ditto
.ElectrodeId Electrode identification 8 ASCII char. 3.2.2.42.
.SmplChanger Calibration on a Titrino ON, OFF 3.2.2.43.
.ActPulse Output of a pulse first, all, OFF 3.2.2.44.
.Statistics Statistics
.Status Status of statistics calculation ON, OFF 3.2.2.26.
.MeanN No.of individual determinations 2...20 ditto
.ResTab Result table
.Select original,delete n,delete all ditto
.DelN Deletion of individual results 1...20 ditto