Metrohm 758 KFD Titrino User Manual Page 107

  • Download
  • Add to my manuals
  • Print
  • Page
    / 210
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 106
3.2 Remote control commands
758 KFD Titrino
103
3.2.2.22. Mode.Parameter.TitrPara.Temp –170.0...25.0...500.0
Parameters for SET, KFT: Titration temperature in °C. If a Pt100 or Pt1000 is
connected, the temperature is measured continuously and the parameter
.Temp is updated.
The temperature is used for the temperature correction in pH measurements.
3.2.2.23. Mode.Parameter.TitrPara.TDelta 1...2...999999
Parameter for SET, KFT: Time interval in s for the entry of a measurement
point in the list of measured points.
3.2.2.24. Mode.Parameter.StopCond.VStop.Type abs., rel., OFF
Mode.Parameter.StopCond.VStop.V 0...99.99...9999.99
Mode.Parameter.StopCond.VStop.Factor 0...±999999
Parameters for SET, KFT: Stop volume.
If an absolute
stop volume (abs.) has been selected, the volume in mL is
valid.
A relative
stop volume (rel.) is dispensed as a function of the sample size:
Stop volume in mL = smpl size factor
The factor is valid.
OFF means that the criterion is not monitored.
3.2.2.25. Mode.Parameter.StopCond.FillRate 0.01...150, max.
Parameters for SET, KFT: Filling rate in the titration in mL/min. Max. means
maximum possible filling rate with the Exchange Unit in current use.
3.2.2.26. Mode.Parameter.Statistics.Status ON, OFF
Mode.Parameter.Statistics.MeanN 2...20
Mode.Parameter.Statistics.ResTab.Selected original, delete n,
delete all
Mode.Parameter.Statistics.ResTab.DelN 1...20
Entries for the statistics calculations.
.Status: On/off switching. Requirement for statistics calculations is a valid
assignment, see 3.2.2.9.
.MeanN: Number of individual results for statistics calculations.
.ResTab.Select: Selection of the table for the statistics calculations.
original: Original table. The original table is (again) set up, i.e. any
individual results which have been deleted are reincorporated
in the statistics calculations.
delete n: Single result lines are removed from the statistics calcu-
lation. All results of the corresponding line in the statistics
table are deleted. Specification of the line number in
.ResTab.DelN.
delete all: Clear entire statistics table. The results can not be reac-
tivated.
.ResTab.DelN: Specification of the line number to be deleted.
Page view 106