E197 Sample size is out of limits.
Exit: The error message disappears on next start or on introduction of new
sample size.
E198 Validation interval is expired.
Exit: The error message disappears on next start or clear counter with
&Config.Monitoring.Validation.ClearCount $G.
E199 Service date is reached.
Exit: The error message disappears on next start or change date in
&Config.Monitoring.Service.Date.
E203 No Oven parameters: Oven not (correctly) connected.
Exit: The error message disappears on next start. If you don't wish oven
parameters in your report, select &Mode.Parameter.Presel.Oven "no" in your
method(s).
E205 Calibration interval is expired.
Exit: The error message disappears on next calibration or if you delete the
calibration.
E212 Transmission error from Remote Box. Unknown characters.
Exit: Rectify error and switch Titrino off and on again.
E213 Time-out error from PC keyboard (Remote Box)
Exit: Rectify error and switch Titrino off and on again.
E214 Check Remote Box. Remote Box not (properly) connected but activated in
&Config.Periph.RemoteBox.
Exit: Rectify error and switch Titrino off and on again.
E270 Overload in dosing element: piston is blocked.
Exit: <STOP>, <QUIT>. Rectify fault at the wet part. Mount it again onto the
dos driver and press <CLEAR>. The dosing element will be initialized.
E282 Overload in dosing element: cock is blocked.
Exit: <STOP>, <QUIT>. Rectify fault at the wet part. Mount it again onto the
dos driver and press <CLEAR>. The dosing element will be initialized.