3.2 Remote control commands
758 KFD Titrino
97
"Setup", continuation
.AutoInfo Automatic message for changes 3.2.2.138.
.Status Switch AutoInfo on/off ON, OFF ditto
.P When mains is switched on ON, OFF ditto
.T Titrator infos
.R When "ready" ON, OFF ditto
.G When method started ON, OFF ditto
.GC When start is initiated ON, OFF ditto
.S When stopped ON, OFF ditto
.B Begin of method ON, OFF ditto
.F End of process ON, OFF ditto
.E Error ON, OFF ditto
.H When "hold" ON, OFF ditto
.C Continue after "hold" ON, OFF ditto
.O Conditioning OK ON, OFF ditto
.N Conditioning not OK ON, OFF ditto
.Re Request after start ON, OFF ditto
.Si Silo empty ON, OFF ditto
.M Entry in measuring point list ON, OFF ditto
.EP Entry in EP list ON, OFF ditto
.RC Recalculation of results done ON, OFF ditto
.C Comport infos
.B1 When COM1 sends a report ON, OFF ditto
.R1 When COM1 is ready again ON, OFF ditto
.B2 When COM2 sends a report ON, OFF ditto
.R2 When COM2 is ready again ON, OFF ditto
.I Changing an I/O input ON, OFF ditto
.O Changing an I/O output ON, OFF ditto
.Graphics Changing the curve output
.Grid Grid on curve ON, OFF 3.2.2.139.
.Frame Frame on curve ON, OFF ditto
.Scale Type of depending axis Full, Auto ditto
.Recorder Length of axes
.Right Length of meas value axis 0.2...0.5...1.00 ditto
.Feed Length of paper drive axis 0.01...0.05...1.00 ditto
.PowerOn RESET (power on) $G 3.2.2.140.
.Initialise Set default values $G 3.2.2.141.
.Select Selection of branch ActMeth,Config,Silo,Calib
Assembly,Setup,All ditto
.RamInit Initialization of working mem. $G 3.2.2.142.
.InstrNo Device Identification $G 3.2.2.143.
.Value Input of device identification 8 ASCII characters ditto
Page view 100