3.2 Remote control commands
758 KFD Titrino
97
"Setup", continuation
├ .AutoInfo Automatic message for changes 3.2.2.138.
│ ├ .Status Switch AutoInfo on/off ON, OFF ditto
│ ├ .P When mains is switched on ON, OFF ditto
│ ├ .T Titrator infos
│ │ ├ .R When "ready" ON, OFF ditto
│ │ ├ .G When method started ON, OFF ditto
│ │ ├ .GC When start is initiated ON, OFF ditto
│ │ ├ .S When stopped ON, OFF ditto
│ │ ├ .B Begin of method ON, OFF ditto
│ │ ├ .F End of process ON, OFF ditto
│ │ ├ .E Error ON, OFF ditto
│ │ ├ .H When "hold" ON, OFF ditto
│ │ ├ .C Continue after "hold" ON, OFF ditto
│ │ ├ .O Conditioning OK ON, OFF ditto
│ │ ├ .N Conditioning not OK ON, OFF ditto
│ │ ├ .Re Request after start ON, OFF ditto
│ │ ├ .Si Silo empty ON, OFF ditto
│ │ ├ .M Entry in measuring point list ON, OFF ditto
│ │ ├ .EP Entry in EP list ON, OFF ditto
│ │ ├ .RC Recalculation of results done ON, OFF ditto
│ ├ .C Comport infos
│ │ ├ .B1 When COM1 sends a report ON, OFF ditto
│ │ ├ .R1 When COM1 is ready again ON, OFF ditto
│ │ ├ .B2 When COM2 sends a report ON, OFF ditto
│ │ ├ .R2 When COM2 is ready again ON, OFF ditto
│ ├ .I Changing an I/O input ON, OFF ditto
│ ├ .O Changing an I/O output ON, OFF ditto
│
├ .Graphics Changing the curve output
│ ├ .Grid Grid on curve ON, OFF 3.2.2.139.
│ ├ .Frame Frame on curve ON, OFF ditto
│ ├ .Scale Type of depending axis Full, Auto ditto
│ ├ .Recorder Length of axes
│ │ ├ .Right Length of meas value axis 0.2...0.5...1.00 ditto
│ │ ├ .Feed Length of paper drive axis 0.01...0.05...1.00 ditto
│
├ .PowerOn RESET (power on) $G 3.2.2.140.
├ .Initialise Set default values $G 3.2.2.141.
│ ├ .Select Selection of branch ActMeth,Config,Silo,Calib
│ Assembly,Setup,All ditto
├ .RamInit Initialization of working mem. $G 3.2.2.142.
├ .InstrNo Device Identification $G 3.2.2.143.
├ .Value Input of device identification 8 ASCII characters ditto
Comments to this Manuals