3.2 Remote control commands
758 KFD Titrino
123
3.2.2.121. Assembly.Bur.Prep $G, $S, $H, $C
Starts the function "preparation". The parameters are under the function
&Config.DosPrep, see 3.2.2.84 up to 3.2.2.91.
3.2.2.122. Assembly.Bur.Rates.Forward.Selected digital, analog
Assembly.Bur.Rates.Forward.Digital 0...150, max.
Assembly.Bur.Rates.Reverse.Selected digital, analog
Assembly.Bur.Rates.Reverse.Digital 0...150, max.
Expel and aspirating rate.
Digital or analog control. With digital control, the inputted value applies (in
mL/min). "max." means maximum possible rate with the Exchange Unit in
current use.
Analog means rate control with the analog potentiometer on Titrino.
3.2.2.123. Assembly.Bur.Fill $G, $H, $C
$G starts the 'FILL' mode of the burette function.
3.2.2.124. Assembly.Bur.ModeDis $G, $S, $H, $C
Assembly.Bur.ModeDis.Selected volume, time
Assembly.Bur.ModeDis.V 0.0001...0.1...9999
Assembly.Bur.ModeDis.Time 0.25...1...86400
Assembly.Bur.ModeDis.VStop 0.0001...9999, OFF
Assembly.Bur.ModeDis.AutoFill ON, OFF
Dispensing mode with parameters. The dispensing mode can only be started
and stopped via the RS Control. During a running dosification, no method can
be started at the Titrino.
.Selected: Dispensing of volume increments or during a preset time.
.Volume, .Time: Size of the volume increments or entry of time.
.VStop: Limit volume for the dispensing.
.AutoFill: ON means automatic filling after every dispensing.
3.2.2.125. Assembly.Meas.Status ON, OFF
Assembly.Meas.MeasInput 1, 2, Diff., Ipol, Upol, Temp
Assembly.Meas.Ipol ±127...1...+127
Assembly.Meas.Upol ±1270...400...+1270
Measurement in assembly. The measuring function can only be started via RS
Control. When the measuring function is switched on, no method can be
started at the Titrino.
.Input: Selection of the potentiometric measuring input 1, 2, diff., polar-
ized electrodes or temperature.
.Ipol: Polarization current in µA.
.Upol: Polarization potential in mV, entry in steps of 10 mV.