Metrohm 758 KFD Titrino User Manual Page 119

  • Download
  • Add to my manuals
  • Print
  • Page
    / 210
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 118
3.2 Remote control commands
758 KFD Titrino
115
3.2.2.94. SmplData.ONSilo.Counter.MaxLines read only
SmplData.ONSilo.Counter.FirstLine read only
SmplData.ONSilo.Counter.LastLine read only
Information on silo memory.
.MaxLines: Maximum possible number of silo lines.
.FirstLine: Lowest valid silo line.
.LastLine: Last occupied silo line.
3.2.2.95. SmplData.ONSilo.EditLine.1.Method up to 8 ASCII characters
SmplData.ONSilo.EditLine.1.Id1 up to 8 ASCII characters
SmplData.ONSilo.EditLine.1.Id2 up to 8 ASCII characters
SmplData.ONSilo.EditLine.1.Id3 up to 8 ASCII characters
SmplData.ONSilo.EditLine.1.ValSmpl 6-digits, sign and dec.point
SmplData.ONSilo.EditLine.1.UnitSmpl up to 5 ASCII characters
SmplData.ONSilo.EditLine.1.C24 read only
SmplData.ONSilo.EditLine.1.C25 read only
SmplData.ONSilo.EditLine.1.Mark read only
etc., up to .255
Contents of a silo line.
.Method: Method used to process the sample, from the method memory or
from the card.
.Id: The identifications Id1...Id3 can also be used as sample-specific
calculation constants C21...C23 in formulas.
.UnitSmpl: If "no unit" is desired for the sample size, the blank string must be
entered.
.C24, .C25: Results which have been assigned to C24 and C25.
.Mark: Mark of the silo line: ""=deleted line, "+"=line which is worked
off, "–"= line which is worked off and not valid for silo calcula-
tions (deleted), "/" last worked-off line, where recalculation can
still be done. Silo lines which have been worked off are "read
only".
3.2.2.96. SmplData.ONSilo.DelLine $G
SmplData.ONSilo.DelLine.LineNum 1...255, OFF
Deletion of a silo line. The line # is deleted with &SmplData.ONSilo.DelLine
$G. If a formerly deleted line is edited again, it becomes valid (function "unde-
lete").
3.2.2.97. SmplData.ONSilo.DelAll $G
Deletes the entire silo memory. Must be triggered with $G.
3.2.2.98. SmplData.ONSilo.CycleLines ON, OFF
Silo data cycling.
With "ON", executed lines are copied to the next free silo lines, see page 65.
Exercise caution if you edit the silo memory during the determinations!
Page view 118
1 2 ... 114 115 116 117 118 119 120 121 122 123 124 ... 209 210

Comments to this Manuals

No comments