3.2 Remote control commands
758 KFD Titrino
82
*Parameter Tree part "Parameters for SET"
├ .SET1 Control parameters for EP1
│ ├ .EP Endpoint 1 depends on meas.quant. 3.2.2.12.
│ ├ .UnitEp Unit of endpoint read only ditto
│ ├ .Dyn Dynamics depends on meas.quant. 3.2.2.13.
│ ├ .UnitDyn Unit of dynamics read only ditto
│ ├ .MaxRate Maximum dosing rate 0.01...10...150, max. ditto
│ ├ .MinRate Minimum dosing rate 0.01...25.0...9999 ditto
│ ├ .Stop Titration stop
│ │ ├ .Type Type of stop criterion drift, time 3.2.2.14.
│ │ ├ .Drift Stop drift 1...20...999 ditto
│ │ ├ .Time Switch-off delay time 0...10...999, inf ditto
│ │ ├ .StopT Stop time 0...999999, OFF ditto
├ .SET2 Control parameters for EP2, as for EP1
├ .TitrPara Titration parameters
│ ├ .Direction Titration direction +, –, auto 3.2.2.15.
│ ├ .XPause Waiting time before start volume 0...999999 3.2.2.16.
│ ├ .Start V Start volume
│ │ ├ .Type Type of start volume abs., rel., OFF 3.2.2.17.
│ │ ├ .V Volume for absolute start volume 0...999.99 ditto
│ │ ├ .Factor Factor for relative start volume 0...±999999 ditto
│ │ ├ .Rate Dispensing rate for start volume 0.01...150.0, max. ditto
│ ├ .Pause Waiting time after start volume 0...999999 3.2.2.18.
│ ├ .ExtrT Extraction time 0...999999 3.2.2.19.
│ ├ .DosUnit Selection of the dosing unit internal D0, external D1,
│ │ external D2 3.2.2.20.
│ ├ .MeasInput Measuring input 1, 2, diff. 3.2.2.21.
│ ├ .Ipol Polarization current 0...1...±127 ditto
│ ├ .Upol Polarization voltage 0...400...±1270 ditto
│ ├ .PolElectrTest Test for polarized electrodes ON, OFF ditto
│ ├ .Temp Titration temperature –170.0...25.0...500.0 3.2.2.22.
│ ├ .TDelta Time interv. for meas.acquisition 1...2...999999 3.2.2.23.
│ │
├
.StopCond Stop conditions
│ ├ .VStop Stop volume
│ │ ├ .Type Type of stop volume abs., rel., OFF 3.2.2.24.
│ │ ├ .V Volume for absolute stop volume 0...99.99...9999.99 ditto
│ │ ├ .Factor Factor for relative stop volume 0...±999999 ditto
│ ├ .FillRate Filling rate 0.01...150.0, max. 3.2.2.25.
│ │
├
.Statistics Statistics
│ ├ .Status Status of statistics calculation ON, OFF 3.2.2.26.
│ ├ .MeanN No. of individual determinations 2...20 ditto
│ ├ .ResTab Result table
│ │ ├ .Select original,delete n,delete all ditto
│ │ ├ .DelN Deletion of individual results 1...20 ditto
│ │
├
.Presel Preselections
│ ├ .Cond Conditioning ON, OFF 3.2.2.27.
│ ├ .DriftDisp Display of drift during cond. ON, OFF ditto
│ ├ .DCor Drift correction
│ │ ├ .Type Type of drift acquisition auto, man., OFF ditto
│ │ ├ .Value Drift value for manual drift corr. 0.0...99.9 ditto
│ ├ .IReq Request of Id's after start id1, id1&2, all, OFF 3.2.2. 28.
│ ├ .SReq Request of smpl size after start value, unit, all, OFF ditto
│ ├ .LimSmplSize Limits for sample size 3.2.2. 29.
│ │ ├ .Status Status of limit control ON, OFF ditto
│ │ ├ .LoLim Lower limit 0.0...999 999 ditto
│ │ ├ .UpLim Upper limit 0.0...999 999 ditto
│ ├ .ActPulse Output of a pulse first, all, cond., OFF 3.2.2.30.
Comments to this Manuals