Metrohm 758 KFD Titrino User Manual Page 116

  • Download
  • Add to my manuals
  • Print
  • Page
    / 210
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 115
3.2 Remote control commands
758 KFD Titrino
112
3.2.2.72. Config.Aux.Set $G
Config.Aux.Set.Date YYYY-MM-DD
Config.Aux.Set.Time hh:mm
Date and time.
Input format of the date: Year-month-day, two-digit, enter leading zeros.
Input format for the time: Hours:minutes, two-digit, enter leading zeros.
Date and time have to be set with &Config.Aux.Set $G just after entry of the
value.
3.2.2.73. Config.Aux.RunNo 0...9999
Current sample number.
Set to 0 on power on and initialization. After 9999, counting starts again at 0.
3.2.2.74. Config.Aux.AutoStart 1...9999, OFF
Number of automatic, internal starts.
3.2.2.75. Config.Aux.StartDelay 0...999999
Start delay time in s. During this time, the data of the preceding determination
are retained.
3.2.2.76. Config.Aux.ResDisplay bold, standard
Character set for the result display at the end of the determination.
3.2.2.77. Config.Aux.DevName up to 8 ASCII characters
Name of the instrument for connections with several units. It is advisable to
use only the letters A...Z (ASCII No. 65...90), a...z (ASCII No. 97...122) and
the numbers 0...9 (ASCII No. 48...57) when the function Setup.AutoInfo
(3.2.2.138) is used at the same time.
If a name has been entered, it will be printed out in the result report (full,
short).
3.2.2.78. Config.Aux.Prog read only
Output of the program version.
The Titrino sends "758.0022" on requests with $Q.
Page view 115
1 2 ... 111 112 113 114 115 116 117 118 119 120 121 ... 209 210

Comments to this Manuals

No comments