3.2 Remote control commands
758 KFD Titrino
100
In place of "RSX", a result name may be entered (.TextRS). This name is out-
putted in the report full, short, scalc full and scalc srt. It is used for the result
and the corresponding mean value.
The limit control for results can also be activated. If a result is out of limit, a
message appears in the result report, E196 is sent, and output line L13 can
be set.
3.2.2.6. Mode.Def.SiloCalc.Assign.C24 RSX, EPX, CXX
Mode.Def.SiloCalc.Assign.C25 RSX, EPX, CXX
Mode.Def.SiloCalc.MatchId id1, id1&2, all, OFF
.Assign.C2X: Assignment to store results in the silo as C2X.
.MatchId: Indication which sample identification(s) have to match so that
the results can be combined.
3.2.2.7. Mode.Def.ComVar.C30 RSX, MNX, EPX, CXX
Mode.Def.ComVar.C31
etc., up to .C39
Assignment of common variables.
The values of the common variables are to be found in &Config.ComVar. They
can be viewed and entered there, see 3.2.2.80.
3.2.2.8. Mode.Def.Report.Assign1
Mode.Def.Report.Assign2
SET, MEAS: param, calib, full, short, mplist, curve, scalc full,
scalc srt, calc, ff
KFT: param, full, short, mplist, curve, scalc full, scalc srt, calc, ff
CAL: param, calib, full, short, scalc full, scalc srt, calc, ff
TIP: param, full, short, scalc full, scalc srt, calc, ff
Definition of the report sequence, which is outputted automatically at the end
of the determination. Entries of more than one block have to be separated with
";".
.Assign1: Output to COM1 of the Titrino. Identical for COM2.
3.2.2.9. Mode.Def.Mean.1.Assign RS1, RSX, EPX, CXX
Mode.Def.Mean.2.Assign
etc., up to .9
Assignment of the statistics calculations. Valid assignments are a requirement
for statistics calculations. In addition, the statistics calculation must be swit-
ched on, see 3.2.2.26. Rules for statistics calculations see page 46.
3.2.2.10. Mode.Def.TempVar.C70 RSX, EPX, CXX
etc. up to .C79
Assignment of temporary variables in a submethod for calculations in TIP.