2.13 Sample degasser
■■■■■■■■■■■■■■■■■■■■■■
40
■■■■■■■■
850 Professional IC – AnCat – MCS
2.13 Sample degasser
The sample degasser removes gas bubbles and disolved gases from the
sample. For degassing, the sample flows into a vacuum chamber through
a special fluoropolymer capillary.
Gas bubbles in the sample lead to poor reproducibility, as the quantity of
sample in the sample loop would not always be the same. Samples (con-
taining gas) should therefore be degassed before injection. For this the
sample is sucked through a degasser chamber before injection, whereby
any gas bubbles are automatically removed.
Note
When using the sample degasser, the rinsing time extends by at least 2
minutes.
1
2
3
4
Figure 22
Sample degasser
1
Sample degasser input
2
Sample degasser output
3
PEEK pressure screw, long 6.2744.090
4
Connection capillaries 6.1803.040
Connecting the sample degasser
1
Remove and keep 6.2744.220 thread stoppers from the input and
output of the sample degasser.
2
Connect the end of the 6.1803.040 sample aspirating capillary con-
nected to the injection valve with a long PEEK pressure screw (22-3)
to the output of the sample degasser (22-2).
3
Connect 6.1803.040 connection capillary with a long PEEK pressure
screw (22-3) to the input of the sample degasser (22-1).