5.1 Problems and their solutions
■■■■■■■■■■■■■■■■■■■■■■
100
■■■■■■■■
850 Professional IC – AnCat – MCS
5 Troubleshooting
5.1 Problems and their solutions
Problem Cause Remedy
Marked drop in
pressure
Leak in the system. Check and seal connections (see Chapter 2.5,
page 14).
Marked rise in pres-
sure
6.2821.120 inline filter
blocked.
Replace 6.2821.130 filter (see Chapter 4.6,
page 83).
MSM – blocked. Regenerate the MSM (see Chapter
4.11.3.1, page 88).
Note: 6.2821.180 pump tubing connection
with filter must be used (29-3).
Conductivity detector
blocked.
Shorten capillary ends by several mm (see
Chapter 4.14.1, Page 97).
Rinse detector against the normal flow
direction (see Chapter 4.14.1, Page 97).
Guard column – blocked. Replace guard column (see Chapter 2.21,
page 62).
Separation column –
blocked.
Regenerate separation column (see Chap-
ter 4.15.4, page 98).
Replace separation column (see "Connect-
ing and rinsing the separation column",
page 65).
Note: Samples should always be microfiltered
(see Chapter 4.7, page 85).
Injection valve – valve
blocked.
Have the valve cleaned (by Metrohm service
technicians).
Drift of the baseline Thermal equilibrium not yet
attained.
Condition (see Chapter 2.15, page 44) instru-
ment with the column thermostat switched
on .
Leak in the system. Check and seal connections (see Chapter 2.5,
page 14).