4 Operation and maintenance
850 Professional IC – AnCat – MCS
■■■■■■■■
95
Order
number
Name Material Inner
diameter
Use
6.1826.330 Pump tubing LFL
(orange/white), 3-
stopper
PVC (Tygon) 0.64 mm No special applications.
6.1826.340 Pump tubing LFL
(black/black), 3-stop-
per
PVC (Tygon) 0.76 mm For the sample solution in
inline dialysis.
6.1826.360 Pump tubing LFL
(white/white), 3-stop-
per
PVC (Tygon) 1.02 mm For sample transfer.
6.1826.380 Pump tubing LFL
(gray/gray), 3-stopper
PVC (Tygon) 1.25 mm For inline sample dilution.
6.1826.390 Pump tubing LFL (yel-
low/yellow), 3-stop-
per
PVC (Tygon) 1.37 mm For the sample solution in
inline ultrafiltration.
4.12.2.2 Pump tubing connection with filter
The 6.2821.130 filters (47-2) should be changed every 3 months, more
frequently at higher backpressure.
1
2
3
Figure 47
Pump tubing connection – Changing the filter
1
Tubing olive
2
Filter 6.2821.130
Packaging contains 10 items.
3
Filter housing
Replacing the filter
1
Unscrewing filter screw
Screw the filter screw (47-3) out of the tubing olive (47-1) with
the aid of two 6.2621.000 adjustable wrenches.
2
Replacing the filter
Remove the old filter (47-2) with tweezers.
Place the new filter (47-2) flat in the tubing olive (47-1) with
tweezers.