3.1. Special messages and error messages
765 Dosimat
20
volume <resol.! The volume to be expelled is smaller than the resolution of the
burette with the exchange unit mounted on the Dosimat.
Exit: Change volume to a value which can be expelled with
the exchange unit mounted on the Dosimat
or
mount an exchange unit where the volume can be
expelled.
V-LIM reached! Security volume is reached.
Exit: <FILL>.
V-PIP > V(B) The stored pipetting volume is higher than the burette volume
of the exchange unit mounted on the Dosimat.
Exit: Change volume to a value which can be expelled with
the exchange unit mounted on the Dosimat
or
mount an exchange unit where the volume can be
expelled.