5.3. Connection of a balance
765 Dosimat
56
5.3 Connection of a balance
The balance is connected to the RS232 output (6) of the Dosimat.
Connecting cable:
Sartorius, MP 8, YDO01... 3.980.3380
Mettler AE, Option 011 3.980.3370 + cable by Mettler: Hand switch ME
42500 or foot switch ME 46278
Mettler AT (15 pin) Cable by Mettler: ME 33640
Mettler, AM/PM: Cable by Mettler: ME 33640 + T-Adapter ME47473 +
Hand switch ME 42500 or foot switch ME 46278
Mettler, AB/AG (LC-RS25): 3.980.3370 + cable by Mettler: LC-RS25
(Adjustments 7 5 2)
Mettler, PG: 3.980.3370 + 6.2125.130
AND Cable on demand
Precisa Cable on demand
Note
• Select the correct balance on the Dosimat (special key 5, see page 18).
• Select the baud rate of the Dosimat such that it corresponds with that of the balance
(special key 1, see page 18).
• Select handshake "none" (special key 6, see page 18).
• Weights can be transferred to parameter "s" of the Dosimat with the aid of a transfer
key on the balance:
- In DOS mode with display s = .
- In CNT D mode with display s g (substance size).
• If another instrument is to be connected to socket A of the Dosimat in addition to the
balance, the 6.2125.000 double plug is available.
• The balance must transmit in mode "g".
5.4 Connection of a printer
The printer is connected to the RS232 output (6) of the Dosimat.
Important: Switch Dosimat on first, then printer.
Connecting cable:
Seiko DPU-414 6.2124.060 + 6.2125.130
Citizen iDP562 6.2124.070
Epson LQ, FX with interface #8148 6.2124.070
Epson LX-300 6.2124.070
Printer:
Seiko DPU-414, 230 V 2.141.0014
Seiko DPU-414, 120 V 2.141.0015
Citizen iDP562, 230 V 2.140.0024
Citizen iDP562, 115 V 2.140.0025