4.1 General
765 Dosimat
33
4 Operation via RS232 Interface
4.1 General
The Dosimat offers an extensive remote control. Data transmission occurs via an interface
according to RS 232 C in half duplex procedure.
The syntax of the commands is based on the following principles:
Commands are strings which always begin with a letter.
Only the first three letters are significant, string length is unlimited.
Actual parameters have to be separated by a space from the preceding command.
The input range for numbers is -1E33 ... -1E-37, 0, 1E-37 ... 1E33.
Examples of possible numbers: 3.567
-.5
5.E4
-123.45E-12
Set 'CR' + 'LF' at the end of a complete command.
'CR': carriage return (ASCII sign No. 13)
'LF': line feed (ASCII sign No. 10)
All commands, where the Dosimat has to send data, begin with letter 'Q' (query).
Each string sent by the Dosimat ends with 'CR' + 'LF'.
Set handshake up for every character (asynchronous transfer).
There are several exceptions in order to avoid problems with time:
The commands GO, STOP, FILL, CLEAR, and INFORMATION are so-called 1-byte
commands. They are simply transmitted as 'G', 'S', 'F', 'C', and 'I' . No separating and
terminating signs are necessary.
With command 'I' (Information), the Dosimat sends two information bytes containing
the information READY, CODE, LOCK, LIMIT etc. see 4.2.7.
Several commands can be transmitted "live", i.e. during a running function. All other
commands are recognized only if the Dosimat is in its ready state, see 4.2. If
commands are not recognized, the corresponding error bytes of information byte 2 are
set and may be inquired with command 'I'.