3.1 Special messages and error messages
765 Dosimat
19
3 Error messages, troubleshooting
blinking value The value keyed in is out of the input range.
3.1 Special messages and error messages
cylinder empty! The Dosimat is set to "auto fill off" and one burette volume has
been expelled in mode DOS.
Exit: <FILL>
error 1 Check sum error in PROM.
Remedy: Call Metrohm-Service
error 2 RAM-check: Error in on-chip-RAM.
Remedy: Call Metrohm-Service
error 3 RAM-check: Error in off-chip-RAM.
Remedy: Call Metrohm-Service
error 4 RAM-check: Error in on- and off-Chip-RAM.
Remedy: Call Metrohm-Service
error 5 Check sum error in off-chip-RAM.
Remedy: RAM has to be re-initialized: Switch Dosimat off.
Press <FILL> during switching it on again. Display shows
"RAM init.". Press <GO> . Display shows "RAM init. passed".
<CLEAR> leads to basic program.
Note: Stored user modes will be cleared on re-initializing of the
RAM and standard mode DOS is loaded into the working
memory.
INF In mode DOS, a result has been calculated with s=0 or the
calculated result is larger than the range of numbers which can
be displayed (1 E+39).
Exit: <CLEAR>
NaN (Not a number). In mode DOS, a result has been calculated
with s=0 and f=0.
Exit: <CLEAR>
no exch. unit! Exchange unit is not (properly) mounted.
Exit: Mount exchange unit properly.
Note: The filling rate is set to maximum.
V> XXXX ml In mode CNT D the volume to be dosed is >999.999 mL.
Exit: <CLEAR> and enter new weight.
V< XXXX ml In mode CNT D the volume to be dosed is smaller than the
smallest possible increment which can be dosed with the ex-
change unit mounted.
Exit: <CLEAR> and enter new weight.