1 Introduction
930 Compact IC Flex SeS (2.930.1400)
■■■■■■■■
1
1 Introduction
1.1 Instrument description
The 930 Compact IC Flex is a professional ion chromatograph. It is distin-
guished by:
Its intelligence: All of the functions are monitored, optimized and
documented in an FDA-compatible manner. Intelligent components,
such as iColumns, save important data onto a chip.
Its compact design: It has a small footprint.
Its transparency: All components are easily accessible and located for
simple visibility and can be monitored during operation through a large
window.
Its safety: The wet end and electronics are physically separated, thereby
preventing fluids from coming into contact with the electronics to a
large extent. A leak sensor is integrated into the wet end.
Its environmental compatibility.
Its low noise emissions.
The intelligent MagIC Net™ software
The 930 Compact IC Flex is operated solely using the MagIC Net software.
A USB cable is used to connect the instrument to a computer with
MagIC Net installed. The intelligent software detects the instrument auto-
matically and checks its functionality. The software controls and monitors
the instrument, evaluates the measured data and manages it in a data-
base.
The 930 Compact IC Flex SeS consists of the following modules:
Leak sensor
The leak sensor detects leaking liquid that collects in the instrument's base
tray. Liquid that leaks in the instrument is routed to the base tray using
drainage tubing and detected there.
High-pressure pump
The intelligent and low-pulsation high-pressure pump pumps the eluent
through the system. It is equipped with a chip where its technical specifi-
cations and "life history" (operating hours, service data, etc.) are saved.
Inline filter
Inline filters protect the separation column reliably from potential contami-
nation from the eluent. The small filter pads with 2 µm pore size can be
replaced quickly and easily. They remove particles from the solutions, such
as bacteria and algae.