7 Technical specifications
930 Compact IC Flex SeS (2.930.1400)
■■■■■■■■
93
Safety shutdown
Function Automatic shutdown upon reaching the pressure limit values
Maximum pres-
sure limit
Adjustable from 0.1 - 50 MPa (1 - 500 bar)
The pump is automatically shut down at the first piston stroke
above the maximum limit value
Minimum pres-
sure limit
Adjustable from 0 - 49 MPa (0 - 490 bar)
The shutdown mechanism is inactive at 0 MPa
The shutdown mechanism becomes active 2 minutes after system
start
The pump is automatically shut down after 3 piston strokes below
the minimum pressure limit
7.7 Injection valve
Actuator switch-
ing time
typ. 100 ms
Maximum operat-
ing pressure
35 MPa (350 bar)
Material PEEK
7.8 Suppressor
Resistance to sol-
vents
No restriction
Switching time typ. 100 ms
7.9 Metrohm CO
2
Suppressor (MCS)
Material
Fluoropolymer
Resistance to sol-
vents
No restriction (except PFC)
Underpressure
Working range Microprocessor controlled/stabilized
Time to estab-
lish after start-
ing
< 30 s
Capillary volume 400 µL
Recommended
flow range
0.1 - 1.0 mL/min
Page view 100