3 Installation
930 Compact IC Flex SeS (2.930.1400)
■■■■■■■■
37
3.11.2 Connecting the MCS
The MCS is connected between the suppressor and the detector.
Connecting the MCS
Figure 14 Connecting the MCS
1
Air aspiration capillary
For drawing in air with low CO
2
content (via
the CO
2
adsorption cartridge).
2
Pressure screw, short (6.2744.070)
Installed on the air aspiration capillary.
3
Luer coupling (6.2744.120)
Mounted on the air aspiration capillary with
a pressure screw (6.2744.070).
1
Removing the threaded stoppers
Remove and keep the two threaded stoppers (6.2744.220) from the
input and output of the MCS.
2
Connection from the suppressor
Use a long pressure screw (6.2744.090) to connect the capillary
labeled out to the input of the of the MCS (labeled In).
3
Connection to the detector
Connect the detector input capillary with a long pressure screw
(6.2744.090) to the MCS output (labeled Out).