3.2. Remote control commands
785 DMP Titrino
90
&Info
Object Description Input range Reference
& Root
:
├ Info Current data
: ├ .Report Transmission of formatted reports $G 3.2.2.92.
│ ├ .Select Report type configuration, parameters,
│ smpl data, statistics, silo, calib
│ C-fmla, def, user method, full,
│ short, mplist, curve, derive, comb,
│ adj para, scalc full, scalc srt, calc,
│ act dir, mem card, all, ff ditto
│
├ .CalibrationData pH calibration data $G 3.2.2.93.
│ ├ .Inp1 For measuring input 1
│ │ ├ .pHas Asymmetry pH 0...7.00...±20.00 ditto
│ │ ├ .Slope Slope of electrode 0...1.000...±9.999 ditto
│ │ ├ .Temp Calibration temperature -170.0...25.0...500.0 ditto
│ │ ├ .Date Date of calibration read only ditto
│ │ ├ .ElectrodeId Id of the calibrated electrode read only ditto
│ ├ .Inp2 For measuring input 2, as for input 1
│ ├ .Diff For differential input, as for input 1
│
├ .Checksums Checksums $G 3.2.2.94.
│ ├ .MPList Checksum of meas.point list read only ditto
│ ├ .ActualMethod Checksum of current method read only ditto
│
├ .DetermData Determination data $G 3.2.2.95.
│ ├ .Write Read/write for several nods ON, OFF
│ ├ .ExV Volume of Exchange/Dosing unit read only/read+write ditto
│ ├ .MPList Measuring point list
│ │ ├ .1 Measuring point 1
│ │ │ ├ .Attribute Attribute read only/read+write ditto
│ │ │ ├ .X X coordinate read only/read+write ditto
│ │ │ ├ .Y Y coordinate read only/read+write ditto
│ │ │ ├ .Z1 Z1 coordinate read only/read+write ditto
│ │ │ ├ .Z2 Z2 coordinate read only/read+write ditto
│ │ │ for each measuring point "Info",
│
├ .TitrResults Titration results
│ ├ .RS Calculated results
│ │ ├ .1 1
st
result
│ │ │ ├ .Value Value read only 3.2.2.96.
│ │ ├ up to 9 results
"Info", continuation
│ ├ .EP Endpoint
│ │ ├ .1 1
st
result
│ │ │ ├ .V Value read only ditto
│ │ │ ├ .Meas Measured value read only
│ │ │ ├ .Mark Mark if more than 1 EP per window read only
│ │ ├ up to 9 EP's
│ ├ .Var Variables C4X
│ │ ├ .C40 Start measured value read only/read+write ditto
│ │ ├ .C41 Titration end volume read only/read+write
│ │ ├ .C42 Titration time read only/read+write