3.2 Remote control commands
785 DMP Titrino
119
Assembly.Bur.ModeDis.VStop 0.0001...9999, OFF
Assembly.Bur.ModeDis.AutoFill ON, OFF
Dispensing mode with parameters. The dispensing mode can only be started
and stopped via the RS Control. During a running dosification, no method can
be started at the Titrino.
.Selected: Dispensing of volume increments or during a preset time.
.Volume, .Time: Size of the volume increments or entry of time.
.VStop: Limit volume for the dispensing.
.AutoFill: ON means automatic filling after every dispensing.
3.2.2.112. Assembly.Meas.Status ON, OFF
Assembly.Meas.MeasInput 1, 2, Diff., Ipol, Upol, Temp
Assembly.Meas.Ipol ±127...1...+127
Assembly.Meas.Upol ±1270...400...+1270
Measurement in assembly. The measuring function can only be started via RS
Control. When the measuring function is switched on, no method can be
started at the Titrino.
.Input: Selection of the potentiometric measuring input 1, 2, diff., polar-
ized electrodes or temperature.
.Ipol: Polarization current in uA.
.Upol: Polarization potential in mV, entry in steps of 10 mV.
3.2.2.113. Assembly.Outputs.AutoEOD ON, OFF
Assembly.Outputs.SetLines $G
Assembly.Outputs.SetLines.L0 active, inactive, pulse, OFF
up to .L13
Assembly.Outputs.ResetLines $G
Setting the I/O output lines.
.AutoEOD: The automatic output of the EOD (End of Determination) at the
end of the determination can be switched off. Thus, for example,
in conjunction with a Titrino several determinations can be per-
formed in the same beaker. Before AutoEOD is switched on, line 3
must be set to "OFF".
.SetLines: With $G, all lines are set.
.SetLines.LX: Set the line LX. "active" means setting of a static signal, "inac-
tive" means resetting of the signal, "pulse" means output of a
pulse of app. 150 ms, "OFF" means the line is not operated, see
also page 166.
Warnings:
If you have "AutoEOD" to "ON", an active line 3 is set to "inac-
tive" by the EOD pulse.
L6 is the line of the activate pulse. An active line 6 is set to "in-
active" by the activate pulse.
L5 is the error line. It is continuously controlled by the Titrino
program and can therefore not be set freely.
Line assignments in Titrino program:
L0 Ready, inactive state
L1 Conditioning OK
L2 Titration in progress
L3 EOD (End Of Determination)