3.2 Remote control commands
785 DMP Titrino
113
short, mplist,curve, deriv, comb, adj para, scalc full,
scalc srt, calc, act dir, mem card, all, ff
$G sends the selected report to the COM which is set in
&Config.PeriphUnit.RepToComport:
configuration: Configuration report. Is not accessible during a running deter-
mination.
parameters: Parameter report of the current method. During a running deter-
mination only "live"-parameters are accessible.
smpl data: Current sample data.
statistics: Statistics table with the individual results.
silo: Contents of the silo memory.
calib: Calibration data of the measuring input in the current method.
C-fmla: Contents of the <C-fmla> key.
def: Contents of the <def> key.
user method: Contents of the method memory.
full: Full result report of the last completed determination.
short: Short result report of the last completed determination.
mplist: Measuring point list of the running determination.
curve: Titration curve of the last determination.
derive: 1st derive of titration curve of the last determination (with DET).
comb: 1st derive combined with the titration curve of the last determina-
tion (with DET).
adj para: Adjustment parameters.
scalc full: Full report of the silo calculations.
scalc srt: Short report of the silo calculations.
calc: Calculation report of the current method.
act dir: Methods of the current directory of the memory card.
mem card: All methods of the memory card.
all: All reports.
ff: Form feed on printer.
Reports which are sent from the Titrino are marked with space (ASCII 32) and
' at the beginning. Then an individual identifier for each report follows. Reports
which are triggered by RS232 ($G) have the same introducer but without pre-
ceding space, i.e. they start with '.
3.2.2.93. Info.CalibrationData $G
Info.CalibrationData.Inp1.pHas -20.00...7.00...+20.00
Info.CalibrationData.Inp1.Slope -9.999...1.000...+9.999
Info.CalibrationData.Inp1.Temp -170.0...25.0...+500.0
Info.CalibrationData.Inp1.Date read only
Info.CalibrationData.Inp1.ElectrodeId read only
identical for .Inp2 and .Diff
pH calibration data for measuring input 1. After the calibration, the data are
entered automatically together with the date of the calibration and the elec-
trode identification, see 3.2.2.44.
Calibration data can be entered. They are accepted with &Info.CalibrationData
$G. If calibration data are entered, the calibration date is deleted.
3.2.2.94. Info.Checksums $G
Info.Checksums.MPList read only
Info.Checksums.ActualMethod read only