3.1 General rules
785 DMP Titrino
75
E130 Wrong sample. For SET with preset titration direction the first
measured value lies behind the endpoint.
Exit: The error message disappears on next startup.
E131 No EP set for SET.
Exit: The error message disappears on next startup.
E132 Silo empty and it has been started with open silo or empty silo has
been opened.
Exit: Send a silo entry.
E133 Silo full.
Exit: Send new command.
E134 No method. A method, which is required from the silo memory or in TIP, does
not exist.
Exit: The error message disappears on next startup.
E135 Check temp.sensor in MEAS T or with activated temperature monitoring.
Exit: Correct error or &m $S.
E136 Same buffer in CAL. Measured value of the second buffer differs less than 6
mV from the measured value of the first buffer.
Exit: Correct error or &m $S.
E137 XXX Bytes are missing so that the method, the silo line could not be stored or
not enough RAM for running TIP.
Exit: Send new command.
E155 No new silo result (C24 or C25).
Exit: The error message disappears on next start or on recalculation.
E157 No sequence defined in TIP.
Exit. The error message disappears on next start.
E158 A second TIP has been called up in TIP.
Exit: The error message disappears on next start.
E160 No new temporary variable.
Exit: The error message disappears on next start.
E161 Measurement range of the secondary measured value (temperature)
exceeded. The primary measured value (pH, U, Ipol, Upol) can also be
unstable.
Exit: Rectify error or &m $S.
E166 Save lines is "OFF" although a submethod of TIP includes an assignment to
C24 or C25.
Exit: The error message disappears on next start. Attention: The data of this
sample will not be stored.
E172 In TIP, a QuickMeas was started, without defining a measuring quantity.
Exit: The error message disappears on next start or &Mode.QuickMeas $S.
E177 Accessing to the memory card, the card was not (properly) inserted.
Exit. The error message disappears on next start.
E178 The date of changing the battery of the card is expired.
Exit. The error message disappears on next start.
E180 Memory card write-protected.
Exit: Send new command.
E181 Memory card not formatted.
Exit: Send new command.
E182 Memory card not accessible.
Exit: Send new command.
E183 A directory with the same name exists already on the memory card.
Exit: Send new command.