Page view 83
3.2. Remote control commands
785 DMP Titrino
80
*Parameter Tree part "Parameters for MET"
.TitrPara Titration parameters
.VStep Volume increment 0...0.10...999.9 3.2.2.12.
.DosRate Dispensing rate for increments 0.01...150.0, max. 3.2.2.13.
.SignalDrift Drift for meas. value acquisition depends on meas.quant. 3.2.2.14.
.UnitSigDrift Unit of measured value drift read only ditto
.EquTime Equilibrium time 0...26...9999, OFF ditto
.StartV Start volume
.Type Type of start volume abs., rel., OFF 3.2.2.15.
.V Volume for absolute start volume 0...999.99 ditto
.Factor Factor for relative start volume 0...±999 999 ditto
.Rate Dispensing rate for start volume 0.01...150.0, max. ditto
.Pause Waiting time 0...999 999 3.2.2.1.
.MeasInput Measuring input 1, 2, diff. 3.2.2.18.
.Ipol Polarization current 0...1...±127 ditto
.Upol Polarization voltage 0...400...±1270 ditto
.PolElectrTest Test for polarized electrodes ON, OFF ditto
.Temp Titration temperature -170.0...25.0...500.0 3.2.2.19.
.StopCond Stop conditions
.VStop Stop volume
.Type Type of stop volume abs., rel., OFF 3.2.2.20.
.V Volume for absolute stop volume 0...99.99...9999.99 ditto
.Factor Factor for relative stop volume 0...±999 999 ditto
.MeasStop Stop measured value pH, U, I depends on meas.quant. 3.2.2.21.
.UnitMStop Unit of stop measured value read only ditto
.EPStop Stop after a number of EP's 1...9, OFF 3.2.2.22.
.FillRate Filling rate 0.01...150.0, max. 3.2.2.23.
.Statistics Statistics
.Status Status of statistics calculation ON, OFF 3.2.2.24.
.MeanN No. of individual determinations 2...20 ditto
.ResTab Result table
.Select original,delete n,delete all ditto
.DelN Deletion of individual results 1...20 ditto
.Evaluation Evaluation
.EPC EP criterion depends on meas.quant. 3.2.2.25.
.Recognition EP recognition
.Select Type of EP recognition all,greatest,last,window,OFF ditto
.Window Window
.1 up to 9 windows
.LowLim Lower limit window 1 depends on meas.quant. ditto
.UpLim Upper limit window 1 depends on meas.quant. ditto
.FixEP Fix endpoints
.1 up to 9 fix EP's
.Value Measured value for fix EP1 depends on meas.quant. ditto
.pK pK or HNP evaluation ON, OFF ditto
.Presel Preselections
.IReq Request of Id's after start id1, id1&2, all, OFF 3.2.2.26.
.SReq Request of sample size after start value, unit, all, OFF ditto
.LimSmplSize Limits for sample size 3.2.2.27
.Status Status of limit control ON, OFF ditto
.LoLim Lower limit 0.0...999 999 ditto
.UpLim Upper limit 0.0...999 999 ditto
.ActPulse Output of a pulse ON, OFF 3.2.2.28.