Metrohm 793 IC Sample Preparation Module User Manual

Browse online or download User Manual for Equipment Metrohm 793 IC Sample Preparation Module. Metrohm 793 IC Sample Preparation Module User Manual

  • Download
  • Add to my manuals
  • Print

Summary of Contents

Page 1 - 793 IC Sample Prep Module

793 IC Sample Prep Module IC Sample Prep Module793Met rohmON ONOFFOFFREMOTEPUMPPOWERONOFFREMOTESTEPACTUATOR Instructions for Use8.793.1003 CH-9

Page 2

1 Introduction 793 IC Sample Prep Module/ 8.793.1003 Instructions for Use 6 1.4 Safety notes 1.4.1 Electrical safety While electrical safety

Page 3 - Table of contents

2.1 Setting up the instrument 793 IC Sample Prep Module/ 8.793.1003 Instructions for Use 7 2 Installation 2.1 Setting up the instrument 2.1.1 Pa

Page 4 - List of figures

2 Installation 793 IC Sample Prep Module/ 8.793.1003 Instructions for Use 8 2.2 Mains connection Follow the instructions below for connecting

Page 5 - 1 Introduction

2.2 Mains connection 793 IC Sample Prep Module/ 8.793.1003 Instructions for Use 9 13131313220 ---- 240 V 100 ---- 120 V 1919191920202020220 ---- 24

Page 6 - 1.2 Parts and controls

2 Installation 793 IC Sample Prep Module/ 8.793.1003 Instructions for Use 10 2.3 Connection to a Modular IC System for Neutralization An importa

Page 7

2.3 Connection to a Modular IC System for Neutralization 793 IC Sample Prep Module/ 8.793.1003 Instructions for Use 11 Please note, that the suppre

Page 8 - 1.3.1 Organization

2 Installation 793 IC Sample Prep Module/ 8.793.1003 Instructions for Use 12 The inlet and outlet capillaries mounted on the SP Module A are then

Page 9 - >PARAM/detector

2.3 Connection to a Modular IC System for Neutralization 793 IC Sample Prep Module/ 8.793.1003 Instructions for Use 13 5 Attach pump tubings x

Page 10 - 1.4 Safety notes

2 Installation 793 IC Sample Prep Module/ 8.793.1003 Instructions for Use 14 6 SP Module A connection 2: Regenerant x Pull the inlet capilla

Page 11 - 2 Installation

2.3 Connection to a Modular IC System for Neutralization 793 IC Sample Prep Module/ 8.793.1003 Instructions for Use 15 8 Startup of 793 IC Sampl

Page 12 - 2.2 Mains connection

CH-9101 Herisau/Switzerland E-Mail [email protected] Internet www.metrohm.com 793 IC Sample Prep Module Instructions for Use

Page 13

2 Installation 793 IC Sample Prep Module/ 8.793.1003 Instructions for Use 16 2.3.3 Settings in the «IC Net 2.1» program During configuration of

Page 14 - Neutralization

2.3 Connection to a Modular IC System for Neutralization 793 IC Sample Prep Module/ 8.793.1003 Instructions for Use 17 start-neutralization.smt 2.

Page 15

2 Installation 793 IC Sample Prep Module/ 8.793.1003 Instructions for Use 18 2.4 Connection to a Modular IC System for Cation Elimination A sec

Page 16

2.42.42.42.4 ConnectionConnectionConnectionConnection totototo aaaa ModularModularModularModular ICICICIC SystemSystemSystemSystem fo

Page 17

2 Installation 793 IC Sample Prep Module/ 8.793.1003 Instructions for Use 20 4 Connect SP Module A to valve A x Screw outlet capillary 292929

Page 18

2.42.42.42.4 ConnectionConnectionConnectionConnection totototo aaaa ModularModularModularModular ICICICIC SystemSystemSystemSystem fo

Page 19

2 Installation 793 IC Sample Prep Module/ 8.793.1003 Instructions for Use 22 2.4.3 Settings in the «IC Net 2.1» program During configuration of

Page 20

3.1 Handling the SP Module A 793 IC Sample Prep Module/ 8.793.1003 Instructions for Use 23 3 Operation 3.1 Handling the SP Module A General The S

Page 21

3 Operation 793 IC Sample Prep Module/ 8.793.1003 Instructions for Use 24 3.2 Manual operation The 793 IC Sample Prep Module cannot be manual

Page 22 - Cation Elimination

3.3 Operation via 732 IC Detector 793 IC Sample Prep Module/ 8.793.1003 Instructions for Use 25 3.3 Operation via 732 IC Detector When operating

Page 23

Table of contents 793 IC Sample Prep Module/ 8.793.1003 Instructions for Use I Table of contents 1 Introduction...

Page 24

3 Operation 793 IC Sample Prep Module/ 8.793.1003 Instructions for Use 26 Switch on: >PROGRAM/edit XXX.X min remote 1✴✴✴✴✴✴✴ Switc

Page 25

3.3 Operation via 732 IC Detector 793 IC Sample Prep Module/ 8.793.1003 Instructions for Use 27 Rotate SP Module A at an event In the 732 IC De

Page 26

4 Malfunctions - Maintenance 793 IC Sample Prep Module/ 8.793.1003 Instructions for Use 28 4 Malfunctions – Maintenance 4.1 Malfunctions and th

Page 27 - 3 Operation

4.1 Malfunctions and their rectification 793 IC Sample Prep Module/ 8.793.1003 Instructions for Use 29 Actuator moves to next position but display

Page 28 - 3.2 Manual operation

4 Malfunctions - Maintenance 793 IC Sample Prep Module/ 8.793.1003 Instructions for Use 30 4.2 Maintenance and servicing 4.2.1 Care The 793 IC S

Page 29 - >SETUP/output

4.2 Maintenance and servicing 793 IC Sample Prep Module/ 8.793.1003 Instructions for Use 31 4.2.4 Exchanging the pump tubing Pump tubings are consu

Page 30

4 Malfunctions - Maintenance 793 IC Sample Prep Module/ 8.793.1003 Instructions for Use 32 4.2.5 Regeneration of the cation exchange material If

Page 31 - ACTUATOR

4.2 Maintenance and servicing 793 IC Sample Prep Module/ 8.793.1003 Instructions for Use 33 4.2.6 Cleaning the SP Module A Cleaning the SP Module A

Page 32

4 Malfunctions - Maintenance 793 IC Sample Prep Module/ 8.793.1003 Instructions for Use 34 40404040393939393838383837373737 Fig. 9: Assembling

Page 33

4.2 Maintenance and servicing 793 IC Sample Prep Module/ 8.793.1003 Instructions for Use 35 4.2.7 Exchanging the cation exchange material Exchangin

Page 34 - 4.2.3 Shutdown

Table of contents 793 IC Sample Prep Module/ 8.793.1003 Instructions for Use II 5 Appendix...

Page 35

4 Malfunctions - Maintenance 793 IC Sample Prep Module/ 8.793.1003 Instructions for Use 36 7 Connect and condition the SP Module A x Reconnec

Page 36

4.2 Maintenance and servicing 793 IC Sample Prep Module/ 8.793.1003 Instructions for Use 37 3 Adjust braking range x Rotate adjustment screw 18

Page 37

5 Appendix 793 IC Sample Prep Module/ 8.793.1003 Instructions for Use 38 5 Appendix 5.1 Technical data Pump Pump type 2-channel peristalti

Page 38

5.1 Technical data 793 IC Sample Prep Module/ 8.793.1003 Instructions for Use 39 Mains connection Mains voltage 115 V: 100...120 V r 10 % 230 V

Page 39

5 Appendix 793 IC Sample Prep Module/ 8.793.1003 Instructions for Use 40 5.2 Standard equipment Subject to changes! All dimensions are given i

Page 40

5.3 Optional accessories 793 IC Sample Prep Module/ 8.793.1003 Instructions for Use 41 2 6.2744.110 PEEK coupling Connection between 6.2821.100

Page 41

5 Appendix 793 IC Sample Prep Module/ 8.793.1003 Instructions for Use 42 5.4 Warranty and conformity 5.4.1 Warranty The warranty on our products

Page 42

5.4 Warranty and conformity 793 IC Sample Prep Module/ 8.793.1003 Instructions for Use 43 5.4.2 EU Declaration of conformity EU Declaration of C

Page 43

5 Appendix 793 IC Sample Prep Module/ 8.793.1003 Instructions for Use 44 5.4.3 Certificate of conformity and system validation Certificate of Co

Page 44 - 5.2 Standard equipment

5.5 Index 793 IC Sample Prep Module / 8.793.1003 Instructions for Use 45 5.5 IndexA Actuator... see SP Module A Actuator hold

Page 45 - 5.3 Optional accessories

1.1 Instrument description 793 IC Sample Prep Module/ 8.793.1003 Instructions for Use 1 1 Introduction 1.1 Instrument description The 793 IC Samp

Page 46 - 5.4 Warranty and conformity

5 Appendix 793 IC Sample Prep Module / 8.793.1003 Instructions for Use 46 Optional accessories... 41 Organic contaminatio

Page 47 - EU Declaration of Conformity

1 Introduction 793 IC Sample Prep Module/ 8.793.1003 Instructions for Use 2 1.2 Parts and controls IC Sample Prep Module793Met rohmON ONOFFOFF

Page 48

1.2 Parts and controls 793 IC Sample Prep Module/ 8.793.1003 Instructions for Use 3 WARNING - Fire Hazard -with the same type and rating of fuseFo

Page 49 - 5.5 Index

1 Introduction 793 IC Sample Prep Module/ 8.793.1003 Instructions for Use 4 Fig. 3: 1.793.0110 SP Module A 1.3 Information about the Instruct

Page 50 - 5 Appendix

1.3 Information about the Instructions for Use 793 IC Sample Prep Module/ 8.793.1003 Instructions for Use 5 1.3.2 Notation and pictograms The fo

Comments to this Manuals