3.1. General rules
785 DMP Titrino
76
E196 Result is out of limits.
Exit: The error message disappears on next start or on recalculation.
E197 Sample size is out of limits.
Exit: The error message disappears on next start or on introduction of new
sample size.
E198 Validation interval is expired.
Exit: The error message disappears on next start or clear counter with
&Config.Monitoring.Validation.ClearCount $G.
E199 Service date is reached.
Exit: The error message disappears on next start or change date in
&Config.Monitoring.Service.Date.
E205 Calibration interval is expired.
Exit: The error message disappears on next calibration or if you delete the
calibration.
E212 Transmission error from Remote Box. Unknown characters.
Exit: Rectify error and switch Titrino off and on again.
E213 Time-out error from PC keyboard (Remote Box)
Exit: Rectify error and switch Titrino off and on again.
E214 Check Remote Box. Remote Box not (properly) connected but activated in
&Config.Periph.RemoteBox.
Exit: Rectify error and switch Titrino off and on again.
E270 Overload in dosing element.
Exit: <STOP> then <CLEAR>. The dosing element will be initialized. If the
error reappears, check the wet part.