3.2 Remote control commands
785 DMP Titrino
101
OFF means that the criterion is not monitored.
3.2.2.22. Mode.Parameter.StopCond.FillRate 0.01...150, max.
Parameters for DET, MET, SET: Filling rate in the titration in mL/min. Max.
means maximum possible filling rate with the Exchange Unit in current use.
3.2.2.23. Mode.Parameter.Statistics.Status ON, OFF
Mode.Parameter.Statistics.MeanN 2...20
Mode.Parameter.Statistics.ResTab.Selected original, delete n,
delete all
Mode.Parameter.Statistics.ResTab.DelN 1...20
Entries for the statistics calculations.
.Status: On/off switching. Requirement for statistics calculations is a valid
assignment, see 3.2.2.9.
.MeanN: Number of individual results for statistics calculations.
.ResTab.Select: Selection of the table for the statistics calculations.
original: Original table. The original table is (again) set up, i.e. any
individual results which have been deleted are reincorporated
in the statistics calculations.
delete n: Single result lines are removed from the statistics calcu-
lation. All results of the corresponding line in the statistics
table are deleted. Specification of the line number in
.ResTab.DelN.
delete all: Clear entire statistics table. The results can not be reac-
tivated.
.ResTab.DelN: Specification of the line number to be deleted.
3.2.2.24. Mode.Parameter.Evaluation.EPC DET: 0...5...200
MET pH: 0.1...0.50...9.99
U, Ipol: 1...30...999
Upol: 0.1...2...99.9
Mode.Parameter.Evaluation.Recognition.Selected all, greatest,
last, window, OFF
Mode.Parameter.Evaluation.Recognition.Window.1.LowLim
pH: 0...±20.00, OFF
U, Ipol: 0...±2000, OFF
Upol: 0...±200.0, OFF
Mode.Parameter.Evaluation.Recognition.Window.1.UpLim
Input range as LowLim
etc. up to 9 windows
Mode.Parameter.Evaluation.FixEP.1.Value pH: 0...±20.00, OFF
U; Ipol: 0...±2000, OFF
etc. up to 9 fix EP's Upol: 0...±200.0, OFF
Mode.Parameter.Evaluation.pK ON, OFF
Parameters for DET and MET: Evaluation of the EP's, see page 20.
.EPC: EP criterion in pH, in mV (with U and Ipol) or in uA (with Upol).
.Recognition.Selected: EP recognition.
all: All endpoints found are recognized.
great: Only the largest EP is recognized.
last: Only the last EP is recognized.