5.4 Diagnosis
730 Sample Changer, Appendix
171
The key code table:
5.4.6 Remote Interface
This diagnosis step tests the functionality of all the output (14) and
input lines (8).
• Prepare instrument for diagnosis (see chapter 5.4.2).
• If necessary, press <ê> several times until
diagnosis
>remote test
• <ENTER>
>remote test
remote test connector ?
• Plug test plug 3.496.8550 into the remote interface without
turning the instrument off.
• <ENTER>
The test runs automatically. If no error occurs, the following appears:
>remote test
remote test ok
• Remove test plug and press <ENTER>.
diagnosis
>RS232 test
Code Key Code Key
1 <HOLD / LEARN> 16 <7 / SAMPLE>
2 <STOP> 17 <4 / PUMP>
3 <START 18 <1 / SCAN>
4 <CONFIG> 19 <0 / DEF>
5 <PARAM> 20 <END>
6 <USER METHOD> 21 <è>
7 22 <CLEAR / RESET>
8 <9 / LIFT> 23 <ENTER>
9 <6 / DOS> 24 <é>
10 <3 / WAIT> 25 <ê>
11 <* / ENDSEQ> 26 <SELECT / TOWER>
12 <8 / MOVE> 27 <QUIT>
13 <5 / STIR> 28 <HOME>
14 <2 / CTRL> 29 <ç>
15 <. / PRINT> 30 <INSERT >
31 <DELETE>
Comments to this Manuals