Metrohm 730 Sample Changer User Manual Page 140

  • Download
  • Add to my manuals
  • Print
  • Page
    / 195
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 139
4.10 Remote Control Commands
730 Sample Changer, Detailed Description
137
:
³
à .Scan Scanning the interfaces $G, $S 4.10.2.56
³Ã .Adress Selection of interface Rm, RS
³ À .Pattern Input signal or data
³ for Rm (Remote): 8 x 1,0 or (bin)
³ ready1, ready2, ready,
³ end1, end2, endmeter
³ for RS (RS232): 14 ASCII characters
³ !.R"
³
à .Ctrl Interface control $G 4.10.2.57
³Ã .Adress Interface selection Rm, RS
³ À .Pattern Output signal or data signal
³ for Rm (Remote): 14 x 1,0 or (bin),
³ START device1, START
³ device2, START device,
³ START dos1, START dos2,
³ START dos, METER mode
³ pH, METER mode T, METER
³ mode U, METER mode I,
³ METER mode C, METER cal
³ pH, METER cal C, METER
³ enter, INIT
³ for RS (RS232): 14 ASCII characters, &M;$G
³
à .Def Re- definitions $G 4.10.2.58
³ Ã .Object Item selection STIRRATE, DOSRATE
³ ³ FILLRATE, LIFTRATE,
³ ³ SHIFTRATE, DRIVE.PORT
³ Ã .Adress Component address dependent upon item
³ À .Value Value dependent upon item
³
à .Wait Waiting time $G, $S 4.10.2.59
³ À .Time Waiting time 0...1...9999 s
³
À .End Changer RESET $G, $S 4.10.2.60
Page view 139