Metrohm 730 Sample Changer User Manual Page 104

  • Download
  • Add to my manuals
  • Print
  • Page
    / 195
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 103
4.4 Changer Commands
730 Sample Changer, Detailed Description
101
The following remote control commands can switch on the Auto-
Info status messages of a Metrohm instrument (for example, in a
start sequence):
&Se.A.T.R"ON" 'Ready' status message
&Se.A.T.F"ON" status message at the end of a
determination
&Se.A.T.S"ON" status message at manual halt
&Se.A.T.G"ON" status message at the start of a method
&Se.A.T.E"ON" status message during an error condition
To be consistent, the corresponding AutoInfo messages should al-
so be switched off again in a final sequence (..."OFF").
Detailed information about the syntax of the Metrohm remote con-
trol language can be found in Chapter 4.9 "Operation via the
RS232 interface" (page 125) or in the instruction manual of your
titration instrument.
Please keep the syntax and conventions of the foreign instruments
or computers the Sample Changer is communicating with.
WAIT
3
WAIT
>sample sequence
11 WAIT 1 s
0…1…9999 s
Waiting time
The WAIT command sets a particular waiting interval during me-
thod processing.
ENDSEQ
>
*
ENDSEQ
>sample sequence
12 ENDSEQ
End of the sequence
End mark of a sequence. This ENDSEQ command can be inser-
ted in any desired command line for test purposes. This has the
effect that the sequence (start, sample, or final sequence) is
processed only up to this line.