4.10 Remote Control Commands
730 Sample Changer, Detailed Description
141
4.10.2.11 Mode.StirRates.1.Rate 1...3...15
etc. until .4
Stirring speed in 15 steps.
4.10.2.12 Mode.DosimatSet.DosUnitNo 1…12
Mode.DosimatSet.1.DosRate 0.01…160 ml/min, max.
Mode.DosimatSet.1.FillRate 0.01…160 ml/min, max.
Mode.DosimatSet.1.DosTube 1...4
Mode.DosimatSet.1.FillTube 1...2...4
Mode.DosimatSet.1.ExchTube 1...2...4
Mode.DosimatSet.1.PrepTube 1...4
Mode.DosimatSet.1.EmptyTube 1...4
etc. until .12
Dosing unit settings.
DosUnitNo: Number of current dosing unit
DosRate: Dosing speed
FillRate: Filling speed
DosTube: Dosing outlet of 700 Dosino
FillTube: Filling inlet of 700 Dosino
ExchTube: Rinsing inlet on exchanging the exchange unit
(see &Assembly.Dos.Value("release") 4.10.2.55)
PrepTube: Dosing outlet on preparing cycle of 700 Dosino
EmptyTube: Air inlet for emptying the 700 Dosino.
4.10.2.13 Mode.ManStop.RemCtl STOP device1, STOP device2,
STOP device∗, 14 Bit (1,0, or ∗)
Mode.ManStop.RSCtl 14 ASCII characters
Manual stop options. Signals or data that are transmitted via the
interfaces after pressing the <STOP> key.