3.15 Mounting the base plate
■■■■■■■■■■■■■■■■■■■■■■
34
■■■■■■■■
855 Robotic Titrosampler
Switch the barcode reader to programming mode and make the following
settings:
1
Select the keyboard layout for the desired country (USA, Germany,
France, Spain, Switzerland (German)). This setting must match the
setting in the control software.
Make sure that the Ctrl characters (ASCII 00 to 31) are allowed to
be sent.
Adjust the settings so that the ASCII character 02 (STX or Ctrl B) is
sent as the first character as "Preamble" or "Prefix Code".
Adjust the settings so that the ASCII character 04 (EOT or Ctrl D) is
sent as the last character as "Postamble" or "Record Suffix" or "Post-
fix Code".
Exit programming mode.
3.15 Mounting the base plate
If it is needed by an application that a determination is not carried out on
the sample rack but in an external measuring cell, a stand plate can be
mounted. It can be placed on the left or on the right of a tower of the 855
Robotic Titrosampler . Any accessories parts can be placed on the support
rod of the stand plate.
Mounting the base plate
6.2001.070
1
2
3
4
5
6
Figure 28 Mounting the base plate